| Customization: | Available |
|---|---|
| After-sales Service: | One Year Warranty, Lifetime Maintenance |
| Warranty: | One Year Warranty, Lifetime Maintenance |
Suppliers with verified business licenses
| Model | UP-5004 |
| Barrel parameters | Inner hole 9.55±0.025mm |
| Pistons parameters | Piston head: 9.475±0.015mm |
| Piston length | H=6.35±0.1mm |
| Paramters | Extrusion hole 1=2.095±0.005mm |
| Temperature parameter | With intelligent temperature control instrument, with four pairs of significant temperature setting control, PID parameters can be automatically set, precision up to ± 0.1 degree centigrade |
| Temperature range | 80degree centigrade~400degree centigrade |
| Temperature accuracy | ±0.2 degree centigrade |
| Display resolution | 0.1degree centigrade |
| Maximum comsuption | < 600W |
| Temperature recovery time | less 4 min. |
| Weight parameters are as follows: | |
| Weight accuracy | ±0.5% |
| Basic configuration | A 0.325kg (including binder bar) |
| B 1.2 kg | |
| C 2.16 kg | |
| D 3.8 kg | |
| E 5.0 kg | |
| F 10 kg | |
| G 12.5 kg | |
| H 21.6 kg | |
| Position detection | |
| Loop distance from up and down | 30mm |
| Control precision | ± 0.1mm |
| Test flow control | |
| Times of cutting the material | 0~10 times |
| Material cutting interval | 0~999s(set reference Table 2) |
| Control flow reaches the set temperature without volatility | |
| Barrel temperature time | 15 min. |
| Material be installed | 1 min. |
| Material sample temperature recovery time | 4 min. |
| When binder set | 1min |